Low temperature · Handheld

Agri-Therm III 6110L

A low-temperature handheld instrument documented for outdoor surface-temperature measurement without an embedded data logger.

Everest Agri-Therm infrared thermometer

Product Specifications

Agri-Therm III 6110L technical details

The following values are taken from the supplied product specification sheet.

Temperature measurement

Scale range
−40°C to 100°C
Resolution
0.1°C
Accuracy
±0.25°C from 0°C to 50°C; ±0.5°C from −40°C to 0°C and 50°C to 100°C
Measurement modes
Surface temperature, ambient dry-bulb temperature, and temperature differential
Response time
<1 second

Optics and operation

Field of view
Variable from 4° to 20° with Vario-Zooom™ field-variable focus
Spectral pass band
8–14 µm
Sighting
TTL/SLR intra-optical LED light sighting; no laser
Operating distance
2 cm to infinity
Emissivity
Settable from 10% to 99%

Interface and environment

Output
Analog millivolt, −400 mV to 1000 mV at 10.0 mV/°C
Operating environment
−20°C to 65°C; up to 99% relative humidity, non-condensing
Power
Rechargeable nickel-metal hydride batteries; listed for 10 hours of continuous use
Data logging
Optional custom data logger with USB connection

Documented uses

Application areas

  • Crop monitoring
  • Environmental research
  • Outdoor measurement

This instrument is included from the documented Everest product catalog. Contact Lightsense to confirm current availability, specifications, and configuration options.

Continue browsing